Amico Dual Connect Console Outlets

 
Features:

  • Flow rate exceeds the requirements of NFPA and CSA
  • Accepts only gas specific adapters
  • Indexed to eliminate interchangeability of gas services
  • Indexed to eliminate interchangeability of gas services
  • Up to 1” [25mm] wall thickness adjustment
  • 360° swivel inlet pipe for easy installation
  • Outlets are also available for M.R.I use
  • 2 in 1 Outlet design
  • Various combinations using Ohmeda, Chemetron and DISS connections

 

Request a Quote specs-pdf

 
General Specifications:
The medical gas outlet shall be an Amico Alert-1 series.

  • Outlets shall be manufactured with a 6 – 1/2” [165 mm] length type “K” 1/2” [12.7 mm] outside diameter (3/8” nominal) size copper inlet pipe stub, which is silver brazed to the outlet body. The body shall be of 1 – 5/16” [33 mm] diameter, one piece brass construction. For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall be rated at a maximum of 200 psi [1,379 kPa] in the event the primary check valve is removed for maintenance. Outlet bodies shall be gas specific by indexing each gas service to a gas specific dual pin indexing arrangement on the respective identification module.
  • A large color coded front plate shall be used for ease of gas identification and aesthetic appeal.
  • A one piece chromed fascia plate shall frame the outlet. With the back rough-in mounted the outlet shall adjust from 3/8” [10 mm] to 1” [25 mm] variation in wall thickness.
  • The outlets shall be of modular design and include a gas specific 1.6 mm [16 ga]. Steel mounting plate designed to permit on-site ganging of multiple outlets, in any order, on 5” [127 mm] spacing.
  • The latch-valve assembly shall be D.I.S.S. (Diameter Index Safety System) and shall only accept corresponding gas specific adapters.
  • All M.R.I. outlets are manufactured from nonferrous materials.
  • All outlets shall be cleaned and degreased for medical gas service, factory assembled and tested.
  • The outlets are ETL Listed.
  • Amico products comply with NFPA-99.